After BI-RADS 2025: Breast Imaging Reports Are Becoming a Computable Risk Language

Why would a revised reporting handbook require radiology departments, algorithm teams, and data providers to redesign their interfaces at the same time? This article traces the handbook from its 63-page first edition in 1993 to the 2025 update, breaking down changes in terminology, structure, and outcome auditing, and discussing the relationship between multimodal breast imaging, the four-category ABC grading, structured fields, AI label interfaces, and quality control for high-quality datasets.

Healthcare and MedicineLanghui Industry Research · R&D Exchange Notes

In everyday breast imaging, BI-RADS functions more like a grammar than a document: the words used for descriptions, the number of conclusion categories, and the next steps are all written into this grammar. The discussion triggered by the 2025 update extends beyond radiology because it changes three things at once: terminology, data structure, and outcome auditing. For hospitals, it means adjusting report templates and follow-up registries; for algorithm teams, it means realigning label boundaries; and for data providers, it means adding an engineering interface between “compliant reports” and “trainable data.”

1. From a 63-Page Booklet to Industry Infrastructure

The full name of BI-RADS is Breast Imaging Reporting and Data System, developed under the leadership of the American College of Radiology (ACR). The first edition in 1993 was only 63 pages long and contained no images. Its core goal was simple: enable different hospitals and physicians to produce comparable conclusions for the same mammogram. Over the next three decades, it gradually incorporated ultrasound, MRI, and contrast-enhanced mammography (CEM), adding follow-up rules, coding systems, and outcome-monitoring modules. Category 4 was divided into A, B, and C tiers, while Category 6 was defined as biopsy-confirmed malignancy and used for pre-treatment baseline documentation.

The key point in this evolution is not that the handbook grew from 63 pages to several hundred, but that its function changed. In its early stage, it addressed “how to describe”; in its middle stage, “how to grade”; and over the past decade, it has increasingly taken on “how to audit” and “how to exchange data with downstream systems.” Once a terminology system is used for quality control, medical-device clinical evaluation, and insurance-payment discussions, it is no longer merely an internal documentation convention within a department—it becomes industry infrastructure.

For this reason, the 2025 update deserves to be read across disciplines. Radiology cares about classification thresholds and follow-up recommendations; information departments care about how fields are stored and how to interface with DICOM SR and reporting systems; algorithm teams care about whether label definitions are sufficient for lesion-level supervision; and data providers care about whether reports produced under the new handbook can be transformed into reviewable, traceable, and reusable training samples. All four perspectives point to the same question: can this handbook be interpreted reliably by machines?

A schematic of the BI-RADS closed loop from image acquisition, descriptive dictionary, and classification conclusions through follow-up registration to outcome auditing
Fig.1 The judgment chain from image acquisition to outcome auditing: the new handbook links description, classification, follow-up, and auditing into a single traceable path.

2. What the New Version Really Changes Is Not the Letters, but the Judgment Chain

The most visible change is the addition and removal of terminology, but what matters more in practice is the rearrangement of the judgment chain. Previously, descriptions, classifications, and follow-up recommendations were often scattered across the report body, departmental consensus, and follow-up notebooks, held together by physicians’ experience. The new version places greater emphasis on clearly documenting the path from input to conclusion: which findings must be described, which combinations change the classification, and when additional assessment must be performed before a follow-up recommendation can be given.

Consider contrast-enhanced mammography, which provides two types of information: morphologic enhancement and background parenchymal enhancement. The former is closer to the lesion-description logic of MRI, while the latter relates to the glandular background. If a report template contains only one “enhancement” checkbox, it will be impossible to distinguish lesion characteristics from background interference or to train a reliable classification model. The new version’s refinement of such new modalities effectively asks reports to take responsibility for “feature separability.”

Background parenchymal enhancement (BPE) in MRI follows the same logic. It is both an interpretive confounder and a biological signal associated with hormonal status and glandular density. Capturing it as a separate field, rather than embedding it in a conclusion sentence, means that the same report can support both clinical decision-making and cohort research. Such changes may seem minor, but they determine whether data can be reused.

Five quality-control gates for breast imaging datasets: acquisition compliance, privacy governance, annotation consistency, label completeness, and outcome traceability
Fig.2 Five quality-control gates for a high-quality dataset: if any gate is missing, bias will emerge during model evaluation.

3. Multimodal Complementarity: No Single Examination Can Answer Every Question

The division of labor among breast imaging modalities is clear in clinical practice. Digital breast tomosynthesis (DBT) excels at separating overlapping tissue in dense breast tissue; ultrasound is suitable for supplemental screening and lesion-biopsy guidance in dense breasts; MRI is particularly valuable for high-risk screening, preoperative assessment, and monitoring response to neoadjuvant therapy; and CEM offers an functional imaging option between X-ray and MRI for institutions with access to contrast agents.

This complementary relationship imposes a hard constraint on data development: lesions that are “the same” across modalities cannot simply be merged into a single sample. The same lesion may appear as a cluster of calcifications on DBT, a hypoechoic area on ultrasound, and an enhancing mass on MRI, with boundaries, size measurements, and morphologic descriptions that do not correspond one-to-one. If a dataset retains only one modality and one label, cross-modal differences will be treated as noise. If multiple modalities are retained without aligning spatial locations, the model may learn the scanning sequence rather than the pathology.

A more practical problem is inconsistent examination pathways. Opportunistic screening, organized screening, outpatient diagnosis, and preoperative assessment involve very different combinations of modalities, intervals, and follow-up strategies. In data governance, it is more reliable to stratify by “examination pathway” before proceeding to lesion-level annotation, rather than aggregating directly by disease type. This is especially evident in breast imaging, a field characterized by multiple modalities and repeated follow-up.

4. Turning Category 4’s ABC Subcategories and Category 6 into Executable Rules

Category 4 means “suspicious for malignancy” and requires histologic evidence. It is subdivided into 4A, 4B, and 4C, corresponding to different malignancy-probability ranges and levels of management intensity. Category 6 refers to biopsy-confirmed malignancy entering pre-treatment assessment. Clinicians are familiar with these categories, but the problems become apparent when they are encoded: are probability ranges reference values or interpretation thresholds? When the same lesion has features of both 4B and 4C, which rule takes precedence?

An engineering approach is to divide classification into three parts: evidence, rules, and conclusion. The evidence layer records findings and their confidence, such as irregular morphology, spiculated margins, and heterogeneous internal enhancement. The rule layer specifies combination logic and priorities. The conclusion layer outputs the category and recommendation. In this way, when two physicians assign different classifications, it is possible to trace which piece of evidence or rule caused the disagreement, rather than seeing only two numbers.

The system must also handle “indeterminate” findings. In the real world, a considerable proportion of reports are limited by image quality, positioning, implants, or prior surgical changes and cannot support a definitive classification. If a template forces a binary choice, annotators can only choose between errors and arbitrary decisions. Retaining a pending-review status and counting it separately in statistics is a prerequisite for trustworthy data, not a procedural compromise.

5. Structured Reporting: Turning Natural Language into Queryable Fields

Structured reporting is more than converting paragraphs into tables. It first requires a stable field dictionary: field names, value domains, units, required status, and dependencies on other fields. In breast reports, glandular density, lesion location (quadrant and clock position), three-axis size, margins, associated findings, prior comparisons, and follow-up recommendations are all objects that can be structured.

At the data-storage level, three common implementation approaches are custom forms in the reporting system, DICOM Structured Reporting (DICOM SR), and FHIR resource mapping for research use. Each has trade-offs: custom forms can be changed quickly but exchange poorly across hospitals; DICOM SR is tightly bound to images and is suitable for devices and the PACS side; FHIR is better suited to inter-hospital and research-platform workflows. The most practical solution is often layered: use forms at the reporting stage to ensure data entry, convert to DICOM SR for archiving, and extract an analysis table on the research side.

A commonly overlooked detail is the “multiple ways of writing the same field.” Size may be written as “1.2×0.9cm,” “12mm,” or “approximately 1 cm,” while location may be written as “upper outer quadrant” or “upper left.” If normalization is not performed at the point of entry, downstream cleaning costs will multiply. A more reliable approach is to move normalization into the data-entry interface: use selectors instead of free text, fixed units instead of handwritten values, and reserve free text for descriptions that truly cannot be enumerated.

Field CategoryTypical FieldsStructuring BenefitsCommon Pitfalls
Glandular Tissue and BackgroundGlandular density, BPE gradeStratified screening and risk modelingRecording background enhancement as a lesion feature
Lesion DescriptionMorphology, margins, internal features, sizeLesion-level supervision and follow-up comparisonMissing three-axis measurements, inconsistent units
Examination PathwayScreening/diagnosis/preoperative, prior comparisonStratified evaluation by pathwayMissing pathway fields causing mixed cohorts
Conclusions and RecommendationsCategory, Category 4 subcategory, follow-up intervalQuality-control metrics and recall analysisStoring the category without its rationale

6. Outcome Auditing: Writing the Report Completes Only Half the Work

Half of BI-RADS’ value comes from outcome monitoring. Recall rate, positive predictive value (PPV), malignancy detection rate, and follow-up compliance for Category 3 can only be calculated when reports are linked to pathology and follow-up outcomes. Many institutions produce well-formatted reports but lack follow-up registries, so audits are limited to periodic spot checks rather than forming a continuous closed loop.

In engineering terms, this requires an association chain from imaging to pathology: examination number, inpatient number or follow-up identifier, biopsy pathology number, surgical records, recurrence, and outcome. In practice, this chain often breaks at the follow-up stage, especially for biopsies performed outside the institution, care received elsewhere, and patients lost to follow-up. Feasible remedies include fixing a follow-up recommendation field in the report template, establishing follow-up reminders, regularly matching data from the medical-record front page and pathology system, and explicitly marking samples that cannot be matched rather than defaulting them to negative.

Defaulting to negative as “no malignancy” is one of the most dangerous habits in data governance. It makes recall rates look better and teaches models incorrect negative examples. A more rigorous approach is to define an “outcome unknown” status and provide interval estimates in statistics. For model evaluation, samples with unknown outcomes can participate in training but should be down-weighted, or at least reported separately in the validation set.

7. Designing the Interface Between AI and BI-RADS

Breast imaging AI commonly involves three types of tasks: lesion detection and segmentation, benign-malignant classification, and image-based risk prediction. The three task types have completely different label requirements. Detection requires pixel-level or box-level boundaries, classification requires lesion-level conclusions, and risk prediction requires follow-up outcomes over time. Combining all three into a single “BI-RADS classification model” often produces inflated metrics that are unusable in clinical practice.

A more reliable interface design uses layered annotation: the bottom layer contains image and lesion geometry, the middle layer contains BI-RADS descriptors, and the top layer contains categories and outcomes. A model may learn only one layer, or perform multitask learning across layers, but evaluation must be reported by layer. For example, detection tasks are evaluated by lesion-level sensitivity and false positives per case; classification tasks by AUC and calibration curves; and risk tasks by time-dependent AUC and decision curves.

External validation is another threshold. Good performance on data from one institution, one device, and one manufacturer does not guarantee usability across hospitals. Differences in image processing, compression force, positioning standards, and reader habits among equipment manufacturers can all alter feature distributions. A practical approach is to build a multicenter test set covering different manufacturers, different distributions of glandular density, and different examination pathways, while clearly documenting its composition in the report.

8. Five Quality-Control Gates for High-Quality Datasets

The quality of breast imaging data cannot be captured by the word “clear” alone. We divide quality control into five gates. The first is acquisition compliance: confirm that images are complete, positioning is standard, and there are no obstructions or artifacts. The second is privacy governance: remove direct identifiers such as names, examination numbers, and dates, while addressing indirect identifiers such as faces, tattoos, and implant markings. The third is annotation consistency: have multiple readers annotate independently, calculate consistency, and send disagreements to arbitration. The fourth is label completeness: confirm that lesion boundaries, descriptors, and category conclusions are all present. The fifth is outcome traceability: link pathology and follow-up data as far as possible.

Consistency should not be judged by a single κ value alone. Consistency in lesion detection, boundary delineation, and descriptor selection respond differently to κ. Dice or Hausdorff distance is commonly used for boundary delineation, Cohen’s or Fleiss’ κ for descriptors, and a specific disagreement matrix is more appropriate for category conclusions. Mixing metrics from different levels in one report conceals the true weak points.

The arbitration mechanism should also be built into the workflow rather than handled through ad hoc meetings. A recommended three-level process is independent annotation by annotators, review by team leads, and expert arbitration, with the version and rationale of each change retained. For disputed lesions, both “majority opinion” and “expert opinion” labels can be preserved and selectively used during training. This keeps the data usable while preserving evidence for future research.

  • Acquisition compliance: completeness, positioning, artifacts, and traceable device parameters
  • Privacy governance: removal of direct identifiers, treatment of indirect identifiers, and documented authorization scope
  • Annotation consistency: layered consistency evaluation for lesion detection, boundaries, descriptors, and categories
  • Label completeness: geometry, descriptions, conclusions, and follow-up recommendation fields all present
  • Outcome traceability: pathology and follow-up linkage, with unknown outcomes marked separately

9. Hospital Implementation: Advancing Templates, Training, and Follow-Up in Parallel

When many hospitals update report templates, they do only one thing: move the new terminology into drop-down menus. Physicians then find the process cumbersome, bypass the template, and continue using free text. The problem is not the physicians but a mismatch between template design and workflow. An effective approach is to shorten high-frequency pathways: expose only necessary fields for routine screening and expand detailed items for difficult cases; automatically carry forward prior comparisons to reduce duplicate entry; and connect follow-up recommendations with the scheduling system so that “recommend follow-up” becomes an executable action.

Training should not merely explain terminology; it should include disagreement cases. Assign the same batch of cases to physicians of different seniority, collect disagreements, and discuss them point by point against the new rules. This is far more effective than one-way lecturing. For sections added or adjusted in the new version, conduct a targeted small-sample test, record changes in classification before and after, and assess the actual impact on recall rate.

Follow-up is the third line, and also the one most likely to break. Start by adding Category 3 and Category 0 cases to the follow-up list, setting reminder points, and assigning dedicated staff to follow up. Then gradually expand to Category 4 biopsy results and postoperative pathology. Once follow-up data become stable, audit metrics become meaningful and model evaluation gains real labels.

10. What This Means for the Industry: From Standalone Algorithms to Screening Networks

When reports become structured data, the industry opportunity is no longer limited to “building a classification model.” Upstream lies equipment and acquisition standards; midstream lies report structure and quality control; downstream lies screening networks, follow-up management, and real-world evidence. Device manufacturers need to design clinical evaluation programs around the new terminology; insurers and health-management organizations need interpretable risk-stratification evidence; and research institutions need reproducible cohorts. All three depend on the same thing: data that are interpretable at the source.

For data providers, this means the deliverable must evolve. In the past, what was delivered was images and annotations; now it must include data dictionaries, annotation manuals, consistency reports, follow-up linkage documentation, and version records. What buyers really care about is not “how many cases” but “how their labels were produced, how disagreements were handled, and whether the process can be reproduced.” This is also a point we have repeatedly validated in large-scale imaging-data projects: traceability is harder than scale—and more valuable.

It is important to note that the new handbook is not a universal key. It addresses report standardization and outcome monitoring, but not device differences, population differences, or missing follow-up. Any claim that “training according to the new version makes models universally usable across hospitals” should be met with requests for multicenter external-validation evidence. For the industry, a healthier attitude is to treat BI-RADS as a shared language, not a performance promise.

11. Conclusion: Translating Consensus into an Operational System

The significance of BI-RADS 2025 ultimately lies not in the handbook’s thickness but in whether it can be written into systems: report templates contain fields, PACS contains structured archives, follow-up systems contain reminders, datasets contain dictionaries and versions, and model evaluation contains layered metrics. If any link is missing, consensus regresses into individual experience.

In a field such as breast imaging, which attracts intense attention and has strong screening attributes, the benefits of standardization will propagate along the chain: physicians write reports faster and more consistently, patients have better access to follow-up, researchers receive cleaner data, and algorithm evaluation moves closer to the real world. There is no shortcut, but every step can be verified.

Langhui's Engineering Capabilities Along This Path

Since 2016, Langhui Technology has focused on AI training data. In medical imaging, it covers acquisition-standard design, structured annotation, and quality control for multimodal data including CT, MR, PET, breast X-ray, ultrasound, and other modalities. We build databases in layers based on lesion geometry, descriptors, category conclusions, and follow-up outcomes, and deliver data dictionaries, annotation manuals, consistency reports, and version records so that every sample can be reviewed.

In breast imaging projects, we first align report templates and field definitions with the clinical team, then conduct small-batch trial annotation to assess consistency and sources of disagreement, before moving to large-scale production. Our annotation team includes professionals with medical-imaging backgrounds, key cases are reviewed by experts, and disagreements are handled through a three-level arbitration process.

If your team is working on BI-RADS-related model training, clinical evaluation of medical devices, or screening-cohort development, you can send us your data requirements, modality combinations, and annotation granularity. We will provide actionable recommendations for acquisition and annotation plans.

FAQ

What are the main differences between BI-RADS 2025 and the 2013 fifth edition?

Public discussion has focused on three areas: more detailed descriptions of new modalities such as contrast-enhanced mammography, clarification of some terminology and classification rules, and stronger outcome-auditing and follow-up fields. Specific provisions should be based on the official handbook published by the ACR and the version adopted by your institution. It is not advisable to modify report templates directly based on secondary interpretations.

What is the easiest pitfall for hospitals when upgrading report templates?

Simply placing the new terminology into drop-down menus without adjusting the workflow. Too many fields in high-frequency screening pathways cause the template to be bypassed and free text to return. First stratify pathways, retain only necessary fields for routine screening, expand fields for difficult cases, and provide disagreement-case training in parallel.

How granular should labels be in a breast imaging AI dataset?

It depends on the task. Detection and segmentation require lesion geometry; classification requires lesion-level descriptors and categories; risk prediction requires follow-up outcomes. Use layered annotation and layered evaluation, and avoid covering all task requirements with a single classification label.

How should annotation consistency be evaluated to be considered reliable?

Evaluate metrics by layer: use sensitivity and false positives per case for detection, Dice or Hausdorff distance for boundaries, κ coefficients for descriptors, and a disagreement matrix for category conclusions. A single κ value is insufficient; review disagreement cases as well.

Can institutions without complete follow-up data conduct BI-RADS-related research?

Yes, but the boundaries must be clearly defined. Mark unknown outcomes separately rather than defaulting to negative; report the proportion of missing outcomes in the validation set; prioritize outcome-related analyses on subsets with more complete follow-up; and explain any selection bias.

References & Further Reading
  1. American College of Radiology, official BI-RADS Reporting and Data Systems page https://www.acr.org/Clinical-Resources/Reporting-and-Data-Systems/Bi-Rads
  2. WHO, Breast cancer fact sheet https://www.who.int/news-room/fact-sheets/detail/breast-cancer
  3. DICOM Standard, Structured Reporting https://www.dicomstandard.org/
  4. FDA, Good Machine Learning Practice for Medical Device Development https://www.fda.gov/medical-devices/software-medical-device-samd/good-machine-learning-practice-medical-device-development-guiding-principles
  5. RSNA, Artificial Intelligence in Medical Imaging https://www.rsna.org/artificial-intelligence

FROM INSIGHT TO YOUR PROJECT

Turn Expert Insight into Your Data Solution

Discuss your data needs with Langhui around the technical path and use cases in this article.

Consultation Topic for This Article:After BI-RADS 2025: Breast Imaging Reports Are Becoming a Computable Risk Language

Business Phone137-5502-0164Business Emaillk@langhuiai.com
WeChat QR code of Langhui's business contact, scan to add
WeChat

Scan to add a business contact
Or click the QR code to view the original image

View original WeChat contact card

Submit Your Project Requirements

Our business team will reach out during working hours. Fields marked * are required.

* Provide at least one of phone or email.

Do not enter personal medical records, ID numbers, original contract content, or other sensitive information. The title and link of this article are submitted with your request so our business team can understand the context.